POWER PROBE

Introduction

The INT500 Insulation and Earth-bond Resistance (Continuity) tester is not only a must-have for electricians but with multiple features and wide ranges of automotive, motor testing, electrical installation, etc.

Introduction

The INT500 Insulation and Earth-bond Resistance (Continuity) tester is not only a must-have for electricians but with multiple features and wide ranges of automotive, motor testing, electrical installation, etc.

Features

- Insulation Test with PI/DAR/Timed
- mA-selectable mode
- Auto bi-direction continuity Test
- Visual/acoustic compare function
- Auto AC/DC Voltage Test with LPF
- 1500 records of Store/Recall 

Features

- Insulation Test with PI/DAR/Timed
- mA-selectable mode
- Auto bi-direction continuity Test
- Visual/acoustic compare function
- Auto AC/DC Voltage Test with LPF
- 1500 records of Store/Recall 

Display Count

4000 counts

Overrange display

>OL Reading or >-OL Reading

Conversion Rate

2 times / second

Dimensions (W x H x D)

96mm x207mm x 54mm with holster

Weight

630g including battery.

Power requirement

AA size ALKALINE Battery * 4

Batteries Life:

1.5 AA ALKALINE battery

Installation Category:

IEC/EN 61010-1, IEC/EN 61010-2-034

Compliance to EN 61557:

 

IEC/EN 61557-2

IEC/EN 61557-4

EMC:

EN 61326-1

Pollution degree:

2

Operating Altitude

 

2000m (6562ft)

Operating temperature:

Non-condensing <5°C, 5°C ~ 30°C ( 80% RH),

30°C ~ 40°C ( 75% RH), 40°C ~ 50°C ( 45%RH)

Storage temperature:

-20°C to 60°C , 0 to 80% R.H. (batteries not fitted)

Temperature Coefficient:

0.15 x (Specified accuracy) /°C, < 20°C or > 26°C .

IP Rating:

IP40

Shock vibration:

Random Vibration per MIL-PRFF Class 2

Drop Protection:

4 feet drop to hardwood on concrete floor.

 

CAT

Application Field

II

It is applicable to test and measuring circuits connected

directly to utilization points (socket outlets and similar

points) of the low-voltage MAINS installation.

III

It is applicable to test and measuring circuits connected

to the distribution part of the buildings low-voltage MAINS

installation.

IV

It is applicable to test and measuring circuits connected at

the source of the buildings low-voltage MAINS installation